Device Types
|
The O2C is available in 12 variations. The variations
differ in number of connected probes, the different number of channels,
measuring depths of a probe and the parameters which can be recorded per
channel.
A measurement channel is usually equipped both with a laser-Doppler-measurement-unit
and with a tissue-spectrometer-unit with white light source (O2C). The
following table shows all types with different available extensions.
Parameter | Devices Types | ||
O2C | O2C-OF | O2C-ATS | |
Velocity and Flow | +1 | +1 | -2 |
SO2 and rHb | +1 | -2 | +1 |
Superficial I - 1 Probe | This device is equipped for the connection of one probe which works in one measurement depth. | ||
Superficial II - 2 Probes | This device is equipped for the connection of two probes which work in one measurement depth at different places. (Application e.g. for right-left-comparison or control and experimental measurement place) | ||
Depthselective I - 1 Probe | This device version has a connection for one probe which enables a place- and time-synchronised recording of values in two measurement depths on the surface and in deeper tissues. | ||
Depthselective II - 2 Probes | Here there are two connections for two probes which enable a place- and time-synchronised recording of values in two measurement depths on the surface and in the deeper tissue of two different places. (Application e.g. for right-left-comparison or control and experimental measurement place) |
1 This device type can measure respective
parameter
2 This device type cannot measure respective parameter
Technical Parameters
Manufacturer: LEA Medizintechnik GmbH
O2C
Parameter | |
Physiological effect | Laser emission / White light |
Classification | Laser device class 3 B, protective class I |
Light sources | Laser diode/ Halogen lamp /LED |
Laser parameter | |
Wavelength | NIR |
Mode | continuous wave |
Beam divergence | NA |
Power | < 30 mW |
Spectrometer | |
Detection range | 450-900 nm |
Resolution | 1 nm |
White standard | Yes |
O2C-OF
Parameter | |
Physiological effect | Laser emission |
Classification | Laser device class 3 B, protective class I |
Light sources | Laser diode |
Laser parameter | |
Wavelength | NIR |
Mode | continuous wave |
Beam divergence | NA |
Power | < 30 mW |
O2C-ATS
Parameter | |
Physiological effect | White light |
Classification | Protective class I |
Light source | Halogen lamp / LED |
Spectrometer | |
Detection range | 450-900 nm |
Resolution | 1 nm |
White standard | Yes |
The following parameters are equal in all types:
Weight | 20 kg -25kg |
Dimension | closed: 490 x 335 x 285 mm (B/H/T),open: 490 x335 x 600 mm (B/H/T) |
Power supply | 115-230 V AC, 50-60 Hz |
Power consumption | 250 VA |
Environment causes | |
Operation temperature | 15-30°C |
Storage & transport temperature | 5-45°C |
Atmospheric humidity | 30-75%, not condensed |
Terrain Height | not higher than 2000m msl |