Device Types
www.lea.de

The O2C is available in 12 variations. The variations differ in number of connected probes, the different number of channels, measuring depths of a probe and the parameters which can be recorded per channel.
A measurement channel is usually equipped both with a laser-Doppler-measurement-unit and with a tissue-spectrometer-unit with white light source (O2C). The following table shows all types with different available extensions.
 
Parameter Devices Types
O2C O2C-OF O2C-ATS
Velocity and Flow +1 +1 -2
SO2 and rHb +1 -2 +1
Superficial I - 1 Probe  This device is equipped for the connection of one probe which works in one measurement depth. 
Superficial II - 2 Probes  This device is equipped for the connection of two probes which work in one measurement depth at different places. (Application e.g. for right-left-comparison or control and experimental measurement place) 
Depthselective I - 1 Probe  This device version has a connection for one probe which enables a place- and time-synchronised recording of values in two measurement depths on the surface and in deeper tissues. 
Depthselective II - 2 Probes  Here there are two connections for two probes which enable a place- and time-synchronised recording of values in two measurement depths on the surface and in the deeper tissue of two different places. (Application e.g. for right-left-comparison or control and experimental measurement place) 

1 This device type can measure respective parameter
2 This device type cannot measure respective parameter

Technical Parameters

Manufacturer: LEA Medizintechnik GmbH

O2C
 
Parameter
Physiological effect Laser emission / White light 
Classification Laser device class 3 B, protective class I
Light sources Laser diode/ Halogen lamp /LED
 
Laser parameter
Wavelength NIR
Mode continuous wave
Beam divergence NA
Power < 30 mW
 
Spectrometer
Detection range 450-900 nm
Resolution 1 nm
White standard Yes

O2C-OF
 
Parameter
Physiological effect Laser emission 
Classification Laser device class 3 B, protective class I
Light sources Laser diode
 
Laser parameter
Wavelength NIR
Mode continuous wave
Beam divergence NA
Power < 30 mW
 

O2C-ATS
 
Parameter
Physiological effect White light 
Classification Protective class I
Light source Halogen lamp / LED
 
 
Spectrometer
Detection range 450-900 nm
Resolution 1 nm
White standard Yes

The following parameters are equal in all types:
 
Weight 20 kg -25kg
Dimension closed: 490 x 335 x 285 mm (B/H/T),open: 490 x335 x 600 mm (B/H/T)
 
Power supply 115-230 V AC, 50-60 Hz
Power consumption 250 VA
 
Environment causes
Operation temperature 15-30°C
Storage & transport temperature 5-45°C
Atmospheric humidity 30-75%, not condensed
Terrain Height not higher than 2000m msl